The Dual Switch Valve is used for fluid drainage from the ventricles into the peritoneum, in cases of...
Aesculap, Inc. began a recall of The Dual Switch Valve is used for fluid drainage from the ventricles into the peritoneum, in cases of hydrocephalus on . FDA classified it as Class II and gives the reason as "Aesculap Inc. (AIC (USA)) initiated a recall of Miethke Shunt System, Dual Switch Valve, due to incorrect product labeling on three units. The incorrect product labeling was the wrong expiration date, 2024-09, on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0502-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- The Dual Switch Valve is used for fluid drainage from the ventricles into the peritoneum, in cases of hydrocephalus.
- Product code
- JXG Shunt, Central Nervous System And Components
- Reason for recall
- Aesculap Inc. (AIC (USA)) initiated a recall of Miethke Shunt System, Dual Switch Valve, due to incorrect product labeling on three units. The incorrect product labeling was the wrong expiration date, 2024-09, on the inner labeling of the sterile bag. However, the expiration date on the outer box was correct. No patient injuries were reported as a result of this issue.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- An important correction & removal notification letter, dated October 24, 2014, was sent to the sole consignee who received the affected device. The letter identified the product, problem, and action to be taken.
- Quantity in commerce
- 1
- Distribution
- US Distribution in Kansas only.
- Product codes and lots
- Item Number FV129 Lot Number 4506244166 Serial Number A10150
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0502-2015 | Class II | Terminated | Voluntary: Firm initiated |