Z-0502-2015Class II

The Dual Switch Valve is used for fluid drainage from the ventricles into the peritoneum, in cases of...

Aesculap Inc / initiated 2014-10-24 / Terminated / root cause: labeling and UDI

Aesculap, Inc. began a recall of The Dual Switch Valve is used for fluid drainage from the ventricles into the peritoneum, in cases of hydrocephalus on . FDA classified it as Class II and gives the reason as "Aesculap Inc. (AIC (USA)) initiated a recall of Miethke Shunt System, Dual Switch Valve, due to incorrect product labeling on three units. The incorrect product labeling was the wrong expiration date, 2024-09, on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0502-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
The Dual Switch Valve is used for fluid drainage from the ventricles into the peritoneum, in cases of hydrocephalus.
Product code
JXG Shunt, Central Nervous System And Components
Reason for recall
Aesculap Inc. (AIC (USA)) initiated a recall of Miethke Shunt System, Dual Switch Valve, due to incorrect product labeling on three units. The incorrect product labeling was the wrong expiration date, 2024-09, on the inner labeling of the sterile bag. However, the expiration date on the outer box was correct. No patient injuries were reported as a result of this issue.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
An important correction & removal notification letter, dated October 24, 2014, was sent to the sole consignee who received the affected device. The letter identified the product, problem, and action to be taken.
Quantity in commerce
1
Distribution
US Distribution in Kansas only.
Product codes and lots
Item Number FV129 Lot Number 4506244166 Serial Number A10150

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0502-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0502-2015Class IITerminatedVoluntary: Firm initiated