Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover...
Zimmer, Inc. began a recall of Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2 Polyaxial... on . FDA classified it as Class II and gives the reason as "Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0574-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2 Polyaxial Spinal System, used for posterior lumbar fusion procedures.
- Product code
- KWP Appliance, Fixation, Spinal Interlaminal
- Reason for recall
- Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Consignees were sent via e-mail a Zimmer Spine "Medical Device Recall" letter dated August 26, 2013.The letter described the problem and the product affected by the recall. The letter described the "Required Action" to be taken by the consignees. Requested consignees to complete the Verification Section (Appendix A). For questions they can contact Zimmer Spine Customer Service at 1 (866) 774-6368 or [email protected].
- Quantity in commerce
- 14 devices
- Distribution
- US nationwide distribution.
- Product codes and lots
- Lot: L75066
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0574-2015 | Class II | Terminated | Voluntary: Firm initiated |