Z-0495-2015Class II

PSI Knee System- Prosthesis, knee patellofemorotibial, semi-constrained, cemented, polymer/meal/polymer The...

Zimmer CAS / initiated 2014-11-03 / Terminated / root cause: software

Zimmer CAS began a recall of PSI Knee System- Prosthesis, knee patellofemorotibial, semi-constrained, cemented, polymer/meal/polymer The Zimmer PSI Knee System is indicated as an... on . FDA classified it as Class II and gives the reason as "Internal discovery found that the MRI scans in some PSI knee guides involved an additional rotation adjustments that could potentially lead to misalignment due to a design flaw". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0495-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer CAS
Product
PSI Knee System- Prosthesis, knee patellofemorotibial, semi-constrained, cemented, polymer/meal/polymer The Zimmer PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Internal discovery found that the MRI scans in some PSI knee guides involved an additional rotation adjustments that could potentially lead to misalignment due to a design flaw.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Zimmer CAS, sent an "URGENT: MEDICAL DEVICE NOTIFICATION" dated November 3, 2014 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to read this notice and be familiar with its contents; complete and return the attachment 1 CERTIFICATE OF ACKNOWLEDGEMENT form via fax to Zimmer Inc. at (574) 372-4265 or scan and email to [email protected]; and maintain a copy of the completed attachment 1 for your files. If you have questions concerning this notice, please call 514-861-4074 ext 292 or email: [email protected]
Quantity in commerce
15
Distribution
US Distribution in states of: AZ, CO, IL, OR, and WA.
Product codes and lots
BKL0398L 20-8070-002-01; DHA0645L 20-8070-002-01; DHA0646R 20-8070-002-02; ECO0516L 20-8070-002-01; GHE0651R 20-8070-002-02; KAD0595L 20-8070-002-01; KWI0523R 20-8070-002-02; MHA0483L 20-8070-002-01; MHA0484R 20-8070-002-02; MKE0659R 20-8070-002-02; MMA0728R 20-8070-002-02; SSC0782L 20-8070-001-01; TCA0582L 20-8070-002-01; THO0803L 20-8070-002-01; TMA0760L 20-8070-002-01;  Expiration dates 5/8/14-12/31/15

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0495-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0495-2015Class IITerminatedVoluntary: Firm initiated