GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal...
GE OEC Medical Systems, Inc began a recall of GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize on . FDA classified it as Class II and gives the reason as "mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0840-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8.
- Quantity in commerce
- 61 units installed in US
- Distribution
- Worldwide Distribution
- Product codes and lots
- GE OEC Brivo 715 Prime Serial Numbers: B2S1301 6 B2S1301 8 B2S1302 1 B2S13022 B2S13025 B2S13039 B2S13060 B2S13063 B2S13065 B2S14001 B2S14007 B2S14022 B2S14039 B2S1404 0 B2S1404 5 B2S1406 5 B2S1406 6
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0840-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0841-2015 | Class II | Terminated | Voluntary: Firm initiated |