510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0840-2015Class II

GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal...

GE Healthcare / initiated 2014-09-15 / Terminated / root cause: labeling and UDI

GE OEC Medical Systems, Inc began a recall of GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize on . FDA classified it as Class II and gives the reason as "mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0840-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Product code
OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
Reason for recall
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Root cause tag
labeling and UDI
FDA root cause
Radiation Control for Health and Safety Act
Action
Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8.
Quantity in commerce
61 units installed in US
Distribution
Worldwide Distribution
Product codes and lots
GE OEC Brivo 715 Prime Serial Numbers:  B2S1301 6 B2S1301 8 B2S1302 1 B2S13022 B2S13025 B2S13039 B2S13060 B2S13063 B2S13065 B2S14001 B2S14007 B2S14022 B2S14039 B2S1404 0 B2S1404 5 B2S1406 5 B2S1406 6

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0840-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0840-2015Class IITerminatedVoluntary: Firm initiated
Z-0841-2015Class IITerminatedVoluntary: Firm initiated