GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal...
GE OEC Medical Systems, Inc began a recall of GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize on . FDA classified it as Class II and gives the reason as "mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0841-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8.
- Quantity in commerce
- 61 units installed in US
- Distribution
- Worldwide Distribution
- Product codes and lots
- GE OEC Brivo 865 Plus Serial Numbers: B4S13011 B4S13012 B4S13016 B4S13018 B4S13020 B4S13029 B4S13031 B4S13035 B4S13038 B4S13043 B4S13054 B4S13056 B4S13059- B4S13061 B4S14002- B4S14004 B4S14007 B4S14015 B4S14016 B4S14022 B4S14022 B4S1403 3 B4S1403 4
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0840-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0841-2015 | Class II | Terminated | Voluntary: Firm initiated |