Z-0563-2015Class II

Micro Introducer Kit, 5F Micro Introducer Kit, Ref No. M4005

PFM Medical GmbH / initiated 2014-10-20 / Terminated / root cause: manufacturing process

Pfm Medical Inc began a recall of Micro Introducer Kit, 5F Micro Introducer Kit, Ref No. M4005 Product Usage: These introducers are used for the percutaneous introduction of therapeutic... on . FDA classified it as Class II and gives the reason as "PFM Medical is recalling catheters and other medical devices because they may exceed USP limits for bacterial endotoxin". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0563-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
PFM Medical GmbH
Product
Micro Introducer Kit, 5F Micro Introducer Kit, Ref No. M4005 Product Usage: These introducers are used for the percutaneous introduction of therapeutic devices, such as PICCS. The PFM CT PICC catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, and power injection of contrast media.
Product code
LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Reason for recall
PFM Medical is recalling catheters and other medical devices because they may exceed USP limits for bacterial endotoxin.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
PFm Medical sent an Urgent Device Recall Notification letter dated November 3, 2014 to all their customers who purchased the catheters and other medical devices. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the recall acknowledgement/reply form and fax it to (760) 758-1167 or email it to [email protected]. Customers with any questions are instructed to contact Jessica Greene at (760) 758-8749.
Quantity in commerce
100 units
Distribution
Worldwide Distribution - US Nationwide in the states of MA, NC, UT, PA, TX, OH, LA and worldwide to: Germany and Malaysia.
Product codes and lots
Lot No. 1302-014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0563-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0560-2015Class IITerminatedVoluntary: Firm initiated
Z-0561-2015Class IITerminatedVoluntary: Firm initiated
Z-0562-2015Class IITerminatedVoluntary: Firm initiated
Z-0563-2015Class IITerminatedVoluntary: Firm initiated
Z-0564-2015Class IITerminatedVoluntary: Firm initiated
Z-0565-2015Class IITerminatedVoluntary: Firm initiated
Z-0566-2015Class IITerminatedVoluntary: Firm initiated
Z-0567-2015Class IITerminatedVoluntary: Firm initiated
Z-0568-2015Class IITerminatedVoluntary: Firm initiated