MAS CardioImmune XL Cardiac Marker Control; Model: CAI-XL1; intended for use in the clinical laboratory as an...
Microgenics Corporation began a recall of MAS CardioImmune XL Cardiac Marker Control; Model: CAI-XL1; intended for use in the clinical laboratory as an assayed control serum for monitoring assay... on . FDA classified it as Class II and gives the reason as "MAS CardioImmune XL, level 1 (lot CXL16011), is showing vial-to-vial variation for the analytes Creatine Kinase-MB (CK-MB) and B-Type Natriuretic Peptide-32 (BNP-32)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0501-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microgenics Corporation
- Product
- MAS CardioImmune XL Cardiac Marker Control; Model: CAI-XL1; intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions in specific cardiac marker determinations.
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- MAS CardioImmune XL, level 1 (lot CXL16011), is showing vial-to-vial variation for the analytes Creatine Kinase-MB (CK-MB) and B-Type Natriuretic Peptide-32 (BNP-32).
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- ThermoFisher sent an Urgent Medical Device Field Corrections dated October 13, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Foreign customers were sent letters via overnight mail and e-mail. ACTIONS TO BE TAKEN BY THE CUSTOMER: 1. Determine if you are using or have inventory of MAS Cardiolmmune XL, lot CXL16011. 2. Discontinue use and destroy any remaining inventory of MAS Cardiolmmune XL, lot CXL16011 per your local waste ordinances. 3. Retain a copy of this letter for your laboratory records. 4. If you have forwarded kits of MAS Cardiolmmune XL (lot CXL16011) to another laboratory, please provide a copy of this letter to them. 5. Complete the attached Medical Device Field Correction Response Form and return the form within 5 days to Thermo Fisher Scientific Technical Service as instructed in the form below. 6. Please call technical service at 1-800-232-3342, then press option 2, then option 2 or e-mail [email protected] to arrange immediate product replacement. If you are a distributor of the product, please contact your affected customer base, advise them of the situation, and provide them with a copy of this letter. You should insert your contact information, e-mail and fax numbers in the Medical Device Field Correction Response Form and request that they return the form to you. You should fill out the distributor section of the attached Medical Device Field Correction Response Form and return the form within 5 days to Thermo Fisher Scientific Technical Service as instructed in the form. Firm intends to try to get a response twice more by mail or telephone to insure we reach all customers. Affected product remaining at firm is quarantined and will be destroyed.
- Quantity in commerce
- 2135 vials.
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AR, AZ, CA CO, CT, DE, FL GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY., and the countries of Australia, Canada, China, Germany, Guam, Hong Kong and Taiwan.
- Product codes and lots
- Model No.: CAI-XL1; Lot Number: CXL16011, Exp. Date: 2016-01-31.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0501-2015 | Class II | Terminated | Voluntary: Firm initiated |