Mammomat Inspiration system
Siemens Medical Solutions USA, Inc began a recall of Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals... on . FDA classified it as Class II and gives the reason as "It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0523-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
- Product code
- MUE Full Field Digital, System, X-Ray, Mammographic
- Reason for recall
- It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.
- FDA root cause
- Use error
- Action
- A safety advisory notice, dated October 28, 2014, was sent to users that identified the product, problem, and actions to be taken to prevent risk to patients and users. Siemens technicians werre instructed to contact customers to schedule a field modification.
- Quantity in commerce
- 3
- Distribution
- US Nationwide Distribution in the state of: WV, NY, and FL..
- Product codes and lots
- model number: 10140000
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0523-2015 | Class II | Terminated | Voluntary: Firm initiated |