Z-0523-2015Class II

Mammomat Inspiration system

Siemens Medical Solutions USA, Inc. / initiated 2014-10-28 / Terminated

Siemens Medical Solutions USA, Inc began a recall of Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals... on . FDA classified it as Class II and gives the reason as "It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0523-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.
FDA root cause
Use error
Action
A safety advisory notice, dated October 28, 2014, was sent to users that identified the product, problem, and actions to be taken to prevent risk to patients and users. Siemens technicians werre instructed to contact customers to schedule a field modification.
Quantity in commerce
3
Distribution
US Nationwide Distribution in the state of: WV, NY, and FL..
Product codes and lots
model number: 10140000

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0523-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0523-2015Class IITerminatedVoluntary: Firm initiated