Z-0838-2015Class II

Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro...

Siemens Healthcare Diagnostics Inc. / initiated 2014-11-05 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro diagnostic device intended to duplicate manual... on . FDA classified it as Class II and gives the reason as "Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0838-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Medical Device Correction, dated November 2014, was issued on November 4th, 2014 to inform customers that Siemens Healthcare Diagnostics has confirmed that reagent probe performance on Dimension Vista systems may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server. Instructions for replacing an associated reagent probe were provided to resolve the issue. Customers are to retain the letter for their laboratory records and forward it to anyone who has received the affected product. Customers are also to complete and return the Field Correction Effectiveness Check Form. Customer should direct their questions to their Siemens Customer Care Center - Technical Solutions at 1-800-441-9250 or their local Siemens technical support representative.
Quantity in commerce
1,949
Distribution
Worldwide Distribution -- Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic Korea, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, and United States.
Product codes and lots
Material Nos.: 10284473, 10488224,  10444801, and 10444802

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0838-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0838-2015Class IITerminatedVoluntary: Firm initiated