Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina...
Illumina Inc began a recall of Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform on . FDA classified it as Class III and gives the reason as "Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0850-2015
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Illumina, Inc.
- Product
- Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.
- Product code
- PFS System, Cystic Fibrosis Transmembrane Conductance Regulator Gene, Variant Gene Sequence Detection
- Reason for recall
- Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- A customer notification letter dated 11/12/14 was sent to all customers who purchased the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay. The letter confirms that there is a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument. At certain genomic locations within the region sequenced by the assay, deletions cannot be reported by the software. The letter outlines the potential issues, the results of Illuminia's investigation, and the firm's next steps. Customers' with technical questions or concerns, are instructed to contact their local support team or Illumina Technical Support at [email protected]. Customers with questions regarding the field action, are instructed to contact Illumina Regulatory Affairs at [email protected]. Customers will be contacted by 11/13/14 by phone and a phone script was provided that includes the procedure, talking points, actions required, and next steps.
- Quantity in commerce
- 19 units
- Distribution
- US: MD, NY Non-US: France
- Product codes and lots
- Instrument Serial No. M70106, M00533L, M70114. Lot No. Shipped: K103313, K103461, K103404, K103487.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0850-2015 | Class III | Terminated | Voluntary: Firm initiated |