Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated...

Bausch & Lomb Incorporated / initiated 2014-11-04 / Terminated / root cause: software

Bausch & Lomb Inc began a recall of Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY on . FDA classified it as Class II and gives the reason as "A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0862-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bausch & Lomb Incorporated
Product
Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY
Product code
HQF Laser, Ophthalmic
Reason for recall
A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Bausch & Lomb contacted customers by phone on November 4, 2014. Customers were informed of the conditions required for the device to malfunction. Customers were also notified that a software update to address the situation will be installed in their system within the next few weeks, and that the firm's service team will be scheduling a visit to each facility to install this software update. Customers were also told that in the meantime they may continue to use the device. For further questions please call (585) 338-6839.
Quantity in commerce
106
Distribution
Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.
Product codes and lots
Serial numbers:  SPC01790, SPC01791, SPC01792, SPC01793, SPC01794, SPC01795, SPC01796, SPC01797, SPC01798, SPC01799, SPC01800, SPC01801, SPC01802, SPC01803, SPC01806, SPC01807, SPC01808, SPC01809, SPC01810, SPC01811, SPC01899, SPC01900, SPC01901, SPC01902, SPC01903, SPC01909, SPC01910, SPC01911, SPC01920, SPC01921, SPC01922, SPC01923, SPC01927, SPC01928, SPC01929, SPC01930, SPC01931, SPC01932, SPC01939, SPC01943, SPC01944, SPC01945, SPC01946, SPC01947, SPC01948, SPC01950, SPC01951, SPC01952, SPC01953, SPC01954, SPC01955, SPC01956, SPC01957, SPC01958, SPC01961, SPC01963, SPC01968, SPC01969, SPC01975, SPC01980, SPC01981, SPC01982, SPC01983, SPC01995, SPC01996, SPC01997, SPC02004, SPC02005, SPC02006, SPC02007, SPC02008, SPC02009, SPC02010, SPC02011, SPC02012, SPC02013, SPC02014, SPC02020, SPC02029, SPC02030, SPC02037, SPC02043, SPC02051, SPC02052, SPC02053, SPC02056, SPC02057, SPC02059, SPC02061, SPC02067, SPC02068, SPC02073, SPC02074, SPC02075, SPC02076, SPC02077, SPC02099, SPC02108, SPC02109, SPC02110, SPC02111, SPC02112, SPC02113, SPC02114, SPC02115 and SPC02120

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0862-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0862-2015Class IITerminatedVoluntary: Firm initiated
Z-0863-2015Class IITerminatedVoluntary: Firm initiated
Z-0864-2015Class IITerminatedVoluntary: Firm initiated