510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0801-2015Class II

Turbo Troponin I For in vitro diagnostic use for the quantitative measurement of troponin I in serum...

Siemens Healthcare Diagnostics Inc. / initiated 2014-10-27 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Turbo Troponin I For in vitro diagnostic use for the quantitative measurement of troponin I in serum, heparinized or EDTA plasma, as an aid in the diagnosis of... on . FDA classified it as Class II. As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0801-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Turbo Troponin I For in vitro diagnostic use for the quantitative measurement of troponin I in serum, heparinized or EDTA plasma, as an aid in the diagnosis of acute myocardial infarction (AMI).
Product code
MMI Immunoassay Method, Troponin Subunit
FDA root cause
Under Investigation by firm
Action
The firm, Siemens, sent an "Urgent Medical Device Recall" letter dated October 27, 2014 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to discontinue use of and discard the kit lots listed; review this letter with your Medical Director, complete and return the Field Correction Effectiveness Check attached to this letter within 7 days via fax to: the Customer Care Center at (312) 275-7795; report any complaints or AE associated with the product to your local Siemens Customer Care Center or your local Siemens technical support representative; contact your local Siemens technical support representative to discuss alternative Siemens solutions; retain this letter with your laboratory records, and forward this letter to those who may have received this product. If you have any questions, contact your local Siemens technical support representative or 914-524-3074.
Quantity in commerce
3,622 kits (3,572 domestic and 50 internationally)
Distribution
Worldwide Distribution: US distribution in states of: CO, FL, MN, MT, and NY; and countries of: Algeria, Argentina, Bolivia, Brazil, Bulgaria, Chile, Columbia, Czech Republic, Ecuador, Egypt, Germany, Hungary, India, Italy, Lithuania, Latvia, Pakistan, Poland, Romania, Russian Fed., Serbia, South Africa, Slovakia, Thailand, Tunisia, and Uruguay.
Product codes and lots
Catalog #'s: 1) LSKTI1 and 2) LSKTI1(D); Siemens Material Number (SMN): 1) 10642238 (US) and 2) 10381014 (OUS); Lot Numbers: 319, 321, 322, 323 and D319, D321 and D322.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0801-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0801-2015Class IITerminatedVoluntary: Firm initiated