VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS...
Ortho-Clinical Diagnostics began a recall of VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS System family member, 2) 5600 Integrated System 1)... on . FDA classified it as Class III and gives the reason as "Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1004-2015
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS System family member, 2) 5600 Integrated System 1) VITROS 4600 - in vitro quantitative measurement of a variety of analytes, 2) VITROS 5600 - in vitro quantitative, semi-quantitative, and qualitative measurement of analytes
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
- Root cause tag
- software
- FDA root cause
- Software Manufacturing/Software Deployment
- Action
- Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated September 29, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2. The resolution to this software anomaly will be in a future version of software. Until the new software is installed on their VITROS System, customers were instructed not to use the calibrator barcode labels for VITROS Calibrator Kit 2. Customers were advised that you manually program the calibration. Customers with questions were instructed to contact Customer Technical Services at 1-800-421-3311.
- Quantity in commerce
- 12 units
- Distribution
- Worldwide Distribution - USA including CO, NC, NJ, NY, PA, MI, WV and Internationally to Canada, Japan and Spain.
- Product codes and lots
- System Product Code: 1) 6802445, 2) 6802413
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1004-2015 | Class III | Terminated | Voluntary: Firm initiated |