Z-1004-2015Class III

VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS...

Ortho-Clinical Diagnostics, Inc. / initiated 2014-09-29 / Terminated / root cause: software

Ortho-Clinical Diagnostics began a recall of VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS System family member, 2) 5600 Integrated System 1)... on . FDA classified it as Class III and gives the reason as "Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1004-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS System family member, 2) 5600 Integrated System 1) VITROS 4600 - in vitro quantitative measurement of a variety of analytes, 2) VITROS 5600 - in vitro quantitative, semi-quantitative, and qualitative measurement of analytes
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated September 29, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2. The resolution to this software anomaly will be in a future version of software. Until the new software is installed on their VITROS System, customers were instructed not to use the calibrator barcode labels for VITROS Calibrator Kit 2. Customers were advised that you manually program the calibration. Customers with questions were instructed to contact Customer Technical Services at 1-800-421-3311.
Quantity in commerce
12 units
Distribution
Worldwide Distribution - USA including CO, NC, NJ, NY, PA, MI, WV and Internationally to Canada, Japan and Spain.
Product codes and lots
System Product Code: 1) 6802445, 2) 6802413

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1004-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1004-2015Class IIITerminatedVoluntary: Firm initiated