510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0573-2015Class II

Arrow Nerve Block Component

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2014-11-10 / Terminated / root cause: labeling and UDI

Arrow International Inc began a recall of Arrow Nerve Block Component, Product Code: AB-18040-N: Permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia... on . FDA classified it as Class II and gives the reason as "Arrow is recalling certain lots of the Continuous Nerve Block Component, Product Code: AB-18040-N, due to a labeling inconsistency. The product lidstock incorrectly identifies the needle in the finished good as 17Ga...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0573-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow Nerve Block Component, Product Code: AB-18040-N: Permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.
Product code
CAZ Anesthesia Conduction Kit
Reason for recall
Arrow is recalling certain lots of the Continuous Nerve Block Component, Product Code: AB-18040-N, due to a labeling inconsistency. The product lidstock incorrectly identifies the needle in the finished good as 17Ga rather than the correct 18Ga. The product included in the package is the correct 18Ga size. No injuries or illnesses have been reported.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
An urgent medical device recall notification, dated November 11, 2014, was sent to consignees which identified the product, problem, and action to be taken. Customers with affected stock were asked to immediately discontinue use and quarantine any products with the catalog and lot number. To return product, the Recall Acknowledgement Form should be completed and faxed to 1-855-419-8507 Attn: Customer Service or email it to [email protected]. A customer service representative will be in contact to provide instructions for the return of product. All customers were asked to complete the Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email it to [email protected].
Quantity in commerce
786
Distribution
US Nationwide Distribution.
Product codes and lots
Product Number: AB-18040-N;  Lot Number:  RF2096425,  RF2010131,  RF1057898,  RF0074365,  RF0035689

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0573-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0573-2015Class IITerminatedVoluntary: Firm initiated