Arrow Nerve Block Component
Arrow International Inc began a recall of Arrow Nerve Block Component, Product Code: AB-18040-N: Permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia... on . FDA classified it as Class II and gives the reason as "Arrow is recalling certain lots of the Continuous Nerve Block Component, Product Code: AB-18040-N, due to a labeling inconsistency. The product lidstock incorrectly identifies the needle in the finished good as 17Ga...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0573-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow Nerve Block Component, Product Code: AB-18040-N: Permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.
- Product code
- CAZ Anesthesia Conduction Kit
- Reason for recall
- Arrow is recalling certain lots of the Continuous Nerve Block Component, Product Code: AB-18040-N, due to a labeling inconsistency. The product lidstock incorrectly identifies the needle in the finished good as 17Ga rather than the correct 18Ga. The product included in the package is the correct 18Ga size. No injuries or illnesses have been reported.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- An urgent medical device recall notification, dated November 11, 2014, was sent to consignees which identified the product, problem, and action to be taken. Customers with affected stock were asked to immediately discontinue use and quarantine any products with the catalog and lot number. To return product, the Recall Acknowledgement Form should be completed and faxed to 1-855-419-8507 Attn: Customer Service or email it to [email protected]. A customer service representative will be in contact to provide instructions for the return of product. All customers were asked to complete the Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email it to [email protected].
- Quantity in commerce
- 786
- Distribution
- US Nationwide Distribution.
- Product codes and lots
- Product Number: AB-18040-N; Lot Number: RF2096425, RF2010131, RF1057898, RF0074365, RF0035689
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0573-2015 | Class II | Terminated | Voluntary: Firm initiated |