Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708...
Teleflex Medical began a recall of Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708. Designed for oral and nasal intubation and are... on . FDA classified it as Class II and gives the reason as "Product Labeling; Units from the affected lot indicates that it contains a size 8mm tracheal tube but may contain a size 7.5mm tracheal tube". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0799-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708. Designed for oral and nasal intubation and are indicated for airway management.
- Product code
- BTR Tube, Tracheal (W/Wo Connector)
- Reason for recall
- Product Labeling; Units from the affected lot indicates that it contains a size 8mm tracheal tube but may contain a size 7.5mm tracheal tube.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified by an Urgent Medical Device Recall Notification letter on 11/12/2014. The letter identified the affected product, as well as, the reason for the recall. Customers are instructed to immediately discontinue use of and quarantine any affected product. They are also instructed to complete and fax or e-mail the enclosed Recall Acknowledgement Form regardless of whether they have affected product in stock or not. Upon receipt of the form, a customer service representative will contact customers to provide an RGA number, as well as, instructions for the return of the product. A local sales representative or customer service should be contacted with questions at 1-866-246-6990.
- Quantity in commerce
- 14,000 units
- Distribution
- Worldwide Distribution -- U.S., Belgium, Canada, and Japan.
- Product codes and lots
- Product Code: 5-10116, Lot number: 01A1400211
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0799-2015 | Class II | Terminated | Voluntary: Firm initiated |