510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0799-2015Class II

Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708...

Teleflex Medical / initiated 2014-11-12 / Terminated / root cause: manufacturing process

Teleflex Medical began a recall of Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708. Designed for oral and nasal intubation and are... on . FDA classified it as Class II and gives the reason as "Product Labeling; Units from the affected lot indicates that it contains a size 8mm tracheal tube but may contain a size 7.5mm tracheal tube". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0799-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
Sheridan/CF Cuffed Tracheal Tube, Sterile, Rx only, Teleflex Medical, Research Triangle Park, NC 27708. Designed for oral and nasal intubation and are indicated for airway management.
Product code
BTR Tube, Tracheal (W/Wo Connector)
Reason for recall
Product Labeling; Units from the affected lot indicates that it contains a size 8mm tracheal tube but may contain a size 7.5mm tracheal tube.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by an Urgent Medical Device Recall Notification letter on 11/12/2014. The letter identified the affected product, as well as, the reason for the recall. Customers are instructed to immediately discontinue use of and quarantine any affected product. They are also instructed to complete and fax or e-mail the enclosed Recall Acknowledgement Form regardless of whether they have affected product in stock or not. Upon receipt of the form, a customer service representative will contact customers to provide an RGA number, as well as, instructions for the return of the product. A local sales representative or customer service should be contacted with questions at 1-866-246-6990.
Quantity in commerce
14,000 units
Distribution
Worldwide Distribution -- U.S., Belgium, Canada, and Japan.
Product codes and lots
Product Code: 5-10116, Lot number: 01A1400211

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0799-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0799-2015Class IITerminatedVoluntary: Firm initiated