Z-0836-2015Class II

adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon...

Ion Beam Applications S.A. / initiated 2014-11-07 / Terminated / root cause: software

Ion Beam Applications S.A. began a recall of adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the... on . FDA classified it as Class II and gives the reason as "Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0836-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ion Beam Applications S.A.
Product
adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Product code
LHN System, Radiation Therapy, Charged-Particle, Medical
Reason for recall
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
The firm, IBA, sent a " User Notice" dated November 7, 2014, to the sole affected site-UPHS Customer. The notice describes the product, problem and actions to be taken. The customer was instructed that the KV/kV source acquisition mode has been disabled and shall not be used for patient alignment until system is officially confirmed to be back within tolerances. IBA will change the kV/kV Single Source acquisition workflow configuration file to be linked to the correct geometrical calibration file and will update the kV/kV Single Source calibration file that includes the colinearity parameters, with the correct values. If you need assistance, contact the IBA Site Staff or the Operations manager at +32 10 475 997 or email: [email protected].
Quantity in commerce
1
Distribution
US Distribution in PA only.
Product codes and lots
PAT.109 (US), Treatment room 4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0836-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0836-2015Class IITerminatedVoluntary: Firm initiated