adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon...
Ion Beam Applications S.A. began a recall of adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the... on . FDA classified it as Class II and gives the reason as "Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0836-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ion Beam Applications S.A.
- Product
- adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
- Product code
- LHN System, Radiation Therapy, Charged-Particle, Medical
- Reason for recall
- Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- The firm, IBA, sent a " User Notice" dated November 7, 2014, to the sole affected site-UPHS Customer. The notice describes the product, problem and actions to be taken. The customer was instructed that the KV/kV source acquisition mode has been disabled and shall not be used for patient alignment until system is officially confirmed to be back within tolerances. IBA will change the kV/kV Single Source acquisition workflow configuration file to be linked to the correct geometrical calibration file and will update the kV/kV Single Source calibration file that includes the colinearity parameters, with the correct values. If you need assistance, contact the IBA Site Staff or the Operations manager at +32 10 475 997 or email: [email protected].
- Quantity in commerce
- 1
- Distribution
- US Distribution in PA only.
- Product codes and lots
- PAT.109 (US), Treatment room 4
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0836-2015 | Class II | Terminated | Voluntary: Firm initiated |