Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final...
Stryker Spine began a recall of Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final tighten to 12 Nm the blocker of the XIA 3, Hooks... on . FDA classified it as Class II and gives the reason as "Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0803-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to final tighten to 12 Nm the blocker of the XIA 3, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker Spine sent an Urgent Medical Device Removal Notification/Acknowledgement Form dated 10/6/2014 via Fed Ex. The letter identified the affected product, discussed the product description, product issue, potential hazards, risk mitigations, and actions needed. Customers were asked to read the notice thoroughly and make sure that it is understood. Additionally, they are to complete the enclosed acknowledgement form and fax or e-mail it back.
- Quantity in commerce
- 15 units
- Distribution
- Distributed to GA, IN, MI, NJ, OH, PA, TX & VA.
- Product codes and lots
- Lot number 4279
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0803-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.