DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is...
Mindray DS USA, Inc. dba Mindray North America began a recall of DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 for use with Mindray DPM 6 and DPM 7... on . FDA classified it as Class II and gives the reason as "Mindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0837-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 for use with Mindray DPM 6 and DPM 7 patient monitors . Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Mindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Mindray DS USA, Inc. sent an Urgent Field Corrective Action Letter dated 10/6/2014 to their customers via certified mail with return receipt. The letter identified the affected product, the problem, adverse effects on the patient, and corrective actions. Customers are asked to either verify that the IBP modules are not affected by following the steps provided in the "Procedures for Verifying IBP Channel Connections;" or to discontinue use of the IBP modules until a Mindray Service Representative contacts them and performs the test to verify the function of the IBP modules. Customers are to contact their local Mindray Service Representative at 1-800-288-2121 to inform them of which corrective action will be taken. Questions should be directed to Diane Arpino at 201-995-8407 or [email protected].
- Quantity in commerce
- 455 units
- Distribution
- Distributed to the states of AL, CA, CO, FL, GA, IA, IL, IN, LA. MA, MO, MS, MT, NC, ND, NY, OH, PA, SC, TX, VA, WA and WV.
- Product codes and lots
- DPM 6 - PN: C-047-000199-00 and DPM 7 - PN: C-047-000200-00, IBP Module P/N: 6800-30-50485
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0837-2015 | Class II | Terminated | Voluntary: Firm initiated |