510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0848-2015Class II

VITROS CHEMISTRY PRODUCTS TRIG Slides, REF 133 6544, IVD For in vitro diagnostic use only. VITROS Chemistry...

Ortho-Clinical Diagnostics, Inc. / initiated 2014-11-06 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS CHEMISTRY PRODUCTS TRIG Slides, REF 133 6544, IVD For in vitro diagnostic use only. VITROS Chemistry Products TRIG Slides quantitatively measure... on . FDA classified it as Class II and gives the reason as "Potential for biased results using the specific lot of VITROS CHEMISTRY PRODUCTS TRIG slides". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0848-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS CHEMISTRY PRODUCTS TRIG Slides, REF 133 6544, IVD For in vitro diagnostic use only. VITROS Chemistry Products TRIG Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Product code
CDT Lipase Hydrolysis/Glycerol Kinase Enzyme, Triglycerides
Reason for recall
Potential for biased results using the specific lot of VITROS CHEMISTRY PRODUCTS TRIG slides.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics sent an Urgent Product Correction Notification dated November 6, 2014, via FedEx overnight courierand/or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to all affected customers. The letter identified the product, the problem, and the acton to be taken by the customer. The letter informed the customers of the issue and advise them to discontinue use of the affected lot and to inform their Laboratory Medical Director if they suspect that a previously reported result may have been affected. For additional questions, please contact OCD's Customer Technical Services at 1-800-421-3311. For questions regarding this recall call 908-218-8776.
Quantity in commerce
USA: 380 units, Foreign: 155 units
Distribution
Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Product codes and lots
Lot 0704-3318-4774 (expiry date 01 Dec 2015)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0848-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0848-2015Class IITerminatedVoluntary: Firm initiated