IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011...
Philips Medical Systems (Cleveland) Inc began a recall of IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number... on . FDA classified it as Class II and gives the reason as "Software defect. In certain circumstances, the application may display incorrect measurements of cardiac and aortic anatomy used to select and size the Transcatheter heart valve (THV) to be implanted". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0473-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 System, x-ray, tomography, computed
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Software defect. In certain circumstances, the application may display incorrect measurements of cardiac and aortic anatomy used to select and size the Transcatheter heart valve (THV) to be implanted .
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Consignees were notified of the Field Safety Notice via certified mail.
- Quantity in commerce
- 32
- Distribution
- Worldwide Distribution: US Distribution in states of: FL, NC, and VT; and the countries of Argentina, Austria, Belgium, Canada, Denmark, France, India, Italy, Japan, Netherlands, United Kingdom, and Vietnam.
- Product codes and lots
- IntelliSpace Portal DX/HX/EX: 96027,96029,96025,96030,96031,85479,79935, 98087,85175,96024,106013,450003, 950002, 950003,35009,35030,870056, 550012, 85081, 97069, 45001985333, 35036, and 85382. IntelliSpace Portal DX/HX/EX Upgrade: 950100, 55003, and 81178. IntelliSpace Portal DX/HX/EX Demo: 85451 and 87290. IntelliSpace Portal IX: 50306, 40219, and 40368. IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0473-2015 | Class II | Terminated | Voluntary: Firm initiated |