Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS
Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display... on . FDA classified it as Class II and gives the reason as "Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0576-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo.plaza printer configuration file. Printouts in anatomical size are not correct and may influence diagnostic decisions and/or therapy.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Siemens sent a Safety Advisory notices letter dated November 11, 2014 to affected customers. The letter identified the affected product, problem, and actions to be taken. For questions call 1-888-826-9702.
- Quantity in commerce
- 149
- Distribution
- US Nationwide Distribution
- Product codes and lots
- model numbers: 10592457, 10863171, 10863172, 10863173, 7502029 , 10496279, with multiple serial numbers: 1187, 3182, 3183, 3212, 3206, 3207, 1361, 10870, 12068, 10733, 3052, 2323, 1496, 3246, 2254, 1089, 1213, 14898, 2626, 1624, 2365, 3087, 13504, 11795, 13921, 2971, 12271, 13090, 9594, 2890, 1372, 12789, 3062, 1856, 2500, 13591, 14005, 3017, 3010, 1373, 9792, 13727, 1374, 14292, 2562, 3165, 2098, 2698, 3211, 2769, 1382, 1165, 2310, 13480, 2324, 2751, 2901, 2897, 2981, 3093, 3050, 3085, 14328, 100544, 100566, 100198, 100600, 100259, 100135, 100133, 100031, 100084, 100233, 200308, 200309, 200310, 200311, 200312, 200313, 200314, 200315, 200316, 200317, 200318, 200319, 200320, 200321, 200322, 200323, 200324, 100104, 100105, 100106, 100107, 100108, 100109, 100110, 100111, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100119, 100120, 100245, 100314, 200277, 100139, 100177, 100027, 100601, 100170, 100025, 100174, 100033, 100186, 100189, 100190, 100191, 100192, 100195, 100196, 100306, 100232, 100185, 100596, 100256, 100599, 100193, 100257, 100198, 100226, 100198, 100181, 100029, 100253, 100301, 100132, 100247, 100248, 100139, 100215, 100230, 100146, 100212, 100539.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0576-2015 | Class II | Terminated | Voluntary: Firm initiated |