Z-0576-2015Class II

Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS

Siemens Medical Solutions USA, Inc. / initiated 2014-11-11 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display... on . FDA classified it as Class II and gives the reason as "Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0576-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo.plaza printer configuration file. Printouts in anatomical size are not correct and may influence diagnostic decisions and/or therapy.
Root cause tag
design
FDA root cause
Component design/selection
Action
Siemens sent a Safety Advisory notices letter dated November 11, 2014 to affected customers. The letter identified the affected product, problem, and actions to be taken. For questions call 1-888-826-9702.
Quantity in commerce
149
Distribution
US Nationwide Distribution
Product codes and lots
model numbers: 10592457, 10863171, 10863172, 10863173, 7502029 , 10496279, with multiple serial numbers: 1187, 3182, 3183, 3212, 3206, 3207,  1361, 10870, 12068, 10733, 3052, 2323, 1496, 3246, 2254, 1089, 1213, 14898,  2626, 1624, 2365, 3087, 13504, 11795, 13921, 2971, 12271, 13090, 9594, 2890, 1372, 12789, 3062, 1856, 2500, 13591, 14005, 3017, 3010, 1373, 9792, 13727, 1374, 14292, 2562, 3165, 2098, 2698, 3211, 2769, 1382, 1165, 2310,  13480, 2324, 2751, 2901, 2897, 2981, 3093, 3050, 3085, 14328, 100544, 100566, 100198, 100600, 100259, 100135, 100133, 100031, 100084, 100233,  200308, 200309, 200310, 200311, 200312, 200313, 200314, 200315, 200316, 200317, 200318, 200319, 200320, 200321, 200322, 200323, 200324, 100104, 100105, 100106, 100107, 100108, 100109, 100110, 100111, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100119, 100120, 100245, 100314, 200277, 100139, 100177, 100027, 100601, 100170, 100025, 100174, 100033, 100186, 100189, 100190, 100191, 100192, 100195, 100196, 100306, 100232, 100185, 100596, 100256, 100599, 100193, 100257, 100198, 100226, 100198, 100181, 100029, 100253, 100301, 100132, 100247, 100248, 100139, 100215, 100230, 100146, 100212, 100539.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0576-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0576-2015Class IITerminatedVoluntary: Firm initiated