McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched...
Aircraft Medical Limited began a recall of McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use... on . FDA classified it as Class II and gives the reason as "The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming disconnected during intubation, specifically when the device is being retracted from the patients mouth". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0494-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aircraft Medical Limited
- Product
- McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use, sterile video laryngoscope blade for use with McGRATH MAC Video Laryngoscopes
- Product code
- CCW Laryngoscope, Rigid
- Reason for recall
- The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming disconnected during intubation, specifically when the device is being retracted from the patients mouth.
- FDA root cause
- Component change control
- Action
- Aircraft Medical Limited sent a recall notification letter dated November 4, 2014 to the sole US consignee. The letter included instructions for the consignee to: 1) check its own stock and its customers' stock for the recalled products; 2) arrange for any remaining stock to be removed from use. Customers with any questions can contact Aircraft Medical by e-mail at [email protected].
- Quantity in commerce
- 40,900 blades (US); 16,000 blades (Foreign)
- Distribution
- Worldwide Distribution - US Nationwide in the state of GA and the countries of Austria, China, Denmark, Finland, France, Hong Kong, Israel, Italy, Netherlands, Poland, Saudi Arabia, Singapore, South Korea, Taiwan, and Turkey.
- Product codes and lots
- Product Code: 350-017-000. Lot Numbers and Expiration Dates: Lot 13121202, Expires 2016/12/11; Lot 13122301, Expires 2016/12/25; Lot 13042401, Expires 2017/5/5; Lot 14030501, Expires 2017/3/16; Lot 14031814, Expires 2017/3/26; Lot 14061616, Expires 2017/6/25.