Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX Complete...
Integra LifeSciences Corp. began a recall of Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX... on . FDA classified it as Class II and gives the reason as "Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0831-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247
- Product code
- GWM Device, Monitoring, Intracranial Pressure
- Reason for recall
- Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Integra sent Urgent- Voluntary Medical Device Recall Letters/Recall Acknowledgement and Return Form dated 11/7/2014 to customers and distributors via traceable courier service.
- Quantity in commerce
- 2,506 units
- Distribution
- Nationwide, Foreign
- Product codes and lots
- CC1 Licox¿ PO2 Microprobe, 200 mm 021213, 040414, 070314, 130114, 240214, 270114, 280314, 300614 CC12 Licox¿ PO2 Microprobe, 300 mm 020614, 040814, 050514, 070414, 070714, 220414 CC1.P1 Licox¿ PMO Combined Probe Containing Oxygen and Temperature 050614, 060214, 060314, 100414, 100714, 130214, 130314, 170114, 190614, 210814, 260614, 310714 IM3ST Licox¿ Complete Brain IMC-Probe Kit: 020614, 030314, 110714, 130114, 140314, 160614, 180714, 200114, 280314 IP1.P Licox¿ Single Lumen Bolt CMPS and Temperature Bolt Kit 030714, 060214, 060314, 070814, 080514, 100714, 120614, 130314, 190614, 310714 IP2.P Licox¿ Double Lumen Bolt CMPS, Temperature and ICP Monitoring Kit 030714, 050614, 060214, 060314, 070814, 080514, 090114, 100714, 130314, 140814, 200214, 200414, 260614, 290814 IT2 Licox¿ Tunneling CMPS and Temperature Monitoring Kit 090114, 100414, 190614, 200214, 260614, 310714
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0831-2015 | Class II | Terminated | Voluntary: Firm initiated |