Z-0847-2015Class II

Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system...

Natus Medical Incorporated / initiated 2014-11-19 / Terminated

Natus Medical Incorporated began a recall of Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy on . FDA classified it as Class II and gives the reason as "Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0847-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Natus Medical Incorporated
Product
Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.
Product code
LBI Unit, Neonatal Phototherapy
Reason for recall
Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.
FDA root cause
Other
Action
Natus Medical has begun mailing the Urgent Recall Notification/Field Safety Notice neoBLUE LED Board Replacements letter (DOC-012737B), dated November 5, 2014, to US consignees. Natus Medical also send the Product Advisory neoBLUE LED Board Replacements letter (DOC-0012738), dated October 30, 2014, to International Distribution Partners. Both US and International consignees also received the TECHNICAL BULLETIN LED Board Kit, neoBLUE 2 LED Phototherapy System, p/n: 001840, Treatment Intensity with revised LED board for neoBLUE 2 systems (p/n: 040904 & 040906) [DOC 008353]; and the Label, neoBLUE Intensity/Height (DOC 011561). Consignees are instructed to measure the light intensity of their devices and to document that information of the device using an adhesive label that is provided. The neoBLUE 3 system and its replacement panel sold today are not affected by this issue. If there are any questions regarding this advisory letter or the attached Technical Bulletin, please call Natus Technical Service at 888-496-2887.
Quantity in commerce
1907 total (1413 in US and 494 international)
Distribution
worldwide
Product codes and lots
All distributed Natus neoBLUE 2 systems with LED PCB replacements received after February 16, 2012. For use only with neoBLUE devices manufactured between 2002 and 2007 (neoBLUE2)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0847-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0847-2015Class IITerminatedVoluntary: Firm initiated