Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1065-2015Class II

SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14...

SCC Soft Computer / initiated 2014-11-26 / Terminated / root cause: software

SCC Soft Computer began a recall of SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0 on . FDA classified it as Class II and gives the reason as "Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1065-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SCC Soft Computer
Product
SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database.
Root cause tag
software
FDA root cause
Software design
Action
On 11/26/2014 SCC Soft Computer sent a Correction Communication to their consignees: Corrective Action: Versions 4.3.0.8 through 4.4.0.2 will have a mandatory hotfix created. Due to the amount of versions affected, the range of availability dates will be 11/24/2014 through 12/30/2014. Utility: A utility will be made available to be run on your Live system that will identify cases in which there is a disparity between the diagnosis text in the document and the database. The utility output will be provided for client review and validation.
Quantity in commerce
108 units
Distribution
Distribution US Nationwide and CANADA.
Product codes and lots
Version 4.3.0.8 Release date: 07/31/2009 - Present;  Version 4.3.0.9 Release date: 11/23/2009 - Present;  Version 4.3.0.10 Release date: 09/20/2010- Present;  Version 4.3.0.11 Release date: 11/11/2010- Present;  Version 4.3.0.12 Release date: 02/07/2011- Present; Version 4.3.0.14 Release date: 11/14/2012- Present;  Version 4.3.0.15 Release date: 09/17/2013- Present;  Version 4.4.0.1 Release date: 12/13/2013 - Present;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1065-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1065-2015Class IITerminatedVoluntary: Firm initiated