SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14...
SCC Soft Computer began a recall of SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0 on . FDA classified it as Class II and gives the reason as "Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1065-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 11/26/2014 SCC Soft Computer sent a Correction Communication to their consignees: Corrective Action: Versions 4.3.0.8 through 4.4.0.2 will have a mandatory hotfix created. Due to the amount of versions affected, the range of availability dates will be 11/24/2014 through 12/30/2014. Utility: A utility will be made available to be run on your Live system that will identify cases in which there is a disparity between the diagnosis text in the document and the database. The utility output will be provided for client review and validation.
- Quantity in commerce
- 108 units
- Distribution
- Distribution US Nationwide and CANADA.
- Product codes and lots
- Version 4.3.0.8 Release date: 07/31/2009 - Present; Version 4.3.0.9 Release date: 11/23/2009 - Present; Version 4.3.0.10 Release date: 09/20/2010- Present; Version 4.3.0.11 Release date: 11/11/2010- Present; Version 4.3.0.12 Release date: 02/07/2011- Present; Version 4.3.0.14 Release date: 11/14/2012- Present; Version 4.3.0.15 Release date: 09/17/2013- Present; Version 4.4.0.1 Release date: 12/13/2013 - Present;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1065-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY