510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0899-2015Class II

Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 5mL Luer lock syringe; Reorder Number: 452010

Abbott Medical / initiated 2014-11-21 / Completed / root cause: manufacturing process

Smiths Medical ASD, Inc. began a recall of Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 5mL Luer lock syringe; Reorder Number: 452010 Product Usage: The Needle Pro Edge Safety... on . FDA classified it as Class II and gives the reason as "Needle is not captured in the needle safety sheath in specific lots". As of , the recall status on record is Completed, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0899-2015
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Abbott Medical
Product
Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 5mL Luer lock syringe; Reorder Number: 452010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intended for injection or aspiration of fluids utilizing a luer lock or luer slip syringe.
Product code
FMI Needle, Hypodermic, Single Lumen
Reason for recall
Needle is not captured in the needle safety sheath in specific lots.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Smith Medical sent on 11/24/2014, consignees an Urgent Field Safety Notice via FedEx Overnight. The notification instructs customers to inspect their inventory and complete and return the attached Urgent Field Safety Notice Confirmation Form to coordinate return of affected product to Smiths Medical for replacement/credit. For questions regarding this information, please contact Smiths Medicals Customer Service Department at 800-258-5361.
Quantity in commerce
10,000
Distribution
US Nationwide Distribution in the states of AZ CA FL GA IL IN LA MA MN MO MS MT NC ND NJ NV NY OH OR PA RI TN TX VA WA WI WV
Product codes and lots
Lot Number: 2755713

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0899-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0897-2015Class IICompletedVoluntary: Firm initiated
Z-0898-2015Class IICompletedVoluntary: Firm initiated
Z-0899-2015Class IICompletedVoluntary: Firm initiated