Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and...
GE Healthcare, LLC began a recall of Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for... on . FDA classified it as Class II and gives the reason as "Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0452-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were sent on 9/4/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref#12218 dated September 4, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers are to perform a system functional check as instructed in the letter. If the check fails, they should contact their local GE Healthcare Service Representative. GE Healthcare will correct all systems at no cost to the customer, and a representative will contact customers to arrange for the correction. For questions they can call US 800-437-1171. For other countries they can call their local GE Healthcare Service Representative.
- Quantity in commerce
- 1 - OUS
- Distribution
- Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
- Product codes and lots
- Mfg Lot or Serial # System ID 00000646778BU1 190020RX26
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0451-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0452-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0453-2015 | Class II | Terminated | Voluntary: Firm initiated |