CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human...
Maquet Datascope Corp - Cardiac Assist Division began a recall of CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration... on . FDA classified it as Class II and gives the reason as "Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1062-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Maquet Inc. sent a Product Recall Medical Device Field Correction letter/Field Correction response form, dated 12/16/2014 via Fed Ex. The letter identified the affected product and the reason for the recall. Users were asked to inspect their inventory for the affected units. If customers have any affected product, they are to follow the instructions provided. Additionally, users are to complete and return the attached Field Correction Response Form to acknowledge receipt of the letter and to indicate completion of the steps outlined. Completed forms should be returned by e-mailing a scanned copy to [email protected] or faxing a copy to 1-973-396-3607. Technical questions should be directed to the Technical Support Department at 1-800-777-4222.
- Quantity in commerce
- 838 units-US; 364 units - OUS
- Distribution
- Worldwide Distribution
- Product codes and lots
- Sensation 7 fr.34 cc IAB Demonstration Unit 0684-DC-0433 0684-DO-0433 0684-DC-0469 0684-DO-0469 Sensation 7 Fr. 40 cc IAB Demonstration Unit 0684-DC-0434 0684-DO-0434 0684-DC-0470 0684-DO-0470 Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units 0684-DC-0567 0684-DO-0567 0684-DC-0568 0684-DO-0568 Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units 0684-DC-0575 0684-DO-0575 0684-DC-0576 0684-DO-0576 Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit 0684-DC-0272 0684-DO-0272
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1062-2015 | Class II | Terminated | Voluntary: Firm initiated |