Z-1062-2015Class II

CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human...

Datascope Corp. / initiated 2014-12-16 / Terminated / root cause: design

Maquet Datascope Corp - Cardiac Assist Division began a recall of CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration... on . FDA classified it as Class II and gives the reason as "Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1062-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.
Root cause tag
design
FDA root cause
Device Design
Action
Maquet Inc. sent a Product Recall Medical Device Field Correction letter/Field Correction response form, dated 12/16/2014 via Fed Ex. The letter identified the affected product and the reason for the recall. Users were asked to inspect their inventory for the affected units. If customers have any affected product, they are to follow the instructions provided. Additionally, users are to complete and return the attached Field Correction Response Form to acknowledge receipt of the letter and to indicate completion of the steps outlined. Completed forms should be returned by e-mailing a scanned copy to [email protected] or faxing a copy to 1-973-396-3607. Technical questions should be directed to the Technical Support Department at 1-800-777-4222.
Quantity in commerce
838 units-US; 364 units - OUS
Distribution
Worldwide Distribution
Product codes and lots
Sensation 7 fr.34 cc IAB Demonstration Unit  0684-DC-0433 0684-DO-0433 0684-DC-0469 0684-DO-0469  Sensation 7 Fr. 40 cc IAB Demonstration Unit 0684-DC-0434 0684-DO-0434 0684-DC-0470 0684-DO-0470  Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units  0684-DC-0567 0684-DO-0567 0684-DC-0568 0684-DO-0568  Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units 0684-DC-0575 0684-DO-0575 0684-DC-0576 0684-DO-0576  Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit 0684-DC-0272 0684-DO-0272

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1062-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1062-2015Class IITerminatedVoluntary: Firm initiated