Z-0984-2015Class II

ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV...

Siemens Medical Solutions USA, Inc. / initiated 2014-11-19 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray... on . FDA classified it as Class II and gives the reason as "There may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0984-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
There may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Planned Action: 1. Providing recommendations for user actions in terms of routine dosimetric quality control checks for any IMRT treatment at very high dose rates (Multiple X) as well as for mARC treatments. 2. Software update for the Function Controller #3, which controls parts of the beam generation. This software update enables the correct switching between Pulse Forming Network (PFN) and reduced Pulse Forming Network (iPFN) in pause state, to reduce the dark current during the movement as much as possible. This software will be installed by a Siemens service representative. 3. Siemens will contact the user to make an appointment to implement the modification system at a date and time when the system will be available. These corrections shall be made free of charge. The deployment of the permanent correction shall be completed by July 1, 2015. For further questions please call (610) 219-6300.
Quantity in commerce
3
Distribution
US Distribution to the states of UT, WI, and NY.
Product codes and lots
Model number: 08139789, serial numbers: 5419,  5784,  5830

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0984-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0984-2015Class IITerminatedVoluntary: Firm initiated