510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0900-2015Class II

Osteobond Bone Cement Bone cement monomer is kitted with bone cement powder to create a cement mantle for...

Zimmer GmbH / initiated 2014-12-08 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of Osteobond Bone Cement Bone cement monomer is kitted with bone cement powder to create a cement mantle for orthopedic implants on . FDA classified it as Class II and gives the reason as "Samples of product from Supplier lot 14D0809 and 14D0808 showed little adhesive transfer between the Tyvek lid and cavity when opened for internal bone cement cure and pairing tests. Visually these samples do not meet...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0900-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Osteobond Bone Cement Bone cement monomer is kitted with bone cement powder to create a cement mantle for orthopedic implants.
Product code
LOD Bone Cement
Reason for recall
Samples of product from Supplier lot 14D0809 and 14D0808 showed little adhesive transfer between the Tyvek lid and cavity when opened for internal bone cement cure and pairing tests. Visually these samples do not meet the requirements of ZWI 43.109 and corresponding visual aid 55-0000-310-01.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
"Urgent Medical Device Recall - Lot Specific" notifications will be sent to the affected distributors, surgeons, and hospital risk managers with instructions for returning the affected product. The letters identified the affected product, reason for recall, risks, and responsibilities. Customers are to complete and return the Inventory Return Certification Form along with the affected product. Questions should be directed to the customer call at 1-800-348-2759.
Quantity in commerce
40 units
Distribution
Distributed to MI, KS, and CA.
Product codes and lots
PMA #N17755, decision date 5/17/1976, reclassified 10/14/1999. Class II device; D033511, Item No. 00-1101-012-00; lot 62849791, 14D0809, Manufactured between 9/11/2014-10/6/2014; EXP date: 4/30/2017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0900-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0900-2015Class IITerminatedVoluntary: Firm initiated