510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0853-2015Class II

Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor...

Zimmer GmbH / initiated 2014-12-04 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for... on . FDA classified it as Class II and gives the reason as "The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0853-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Notifications will be sent to the affected distributors and hospital risk managers with instructions for returning the affected product the week of December 8, 2014. Recalled product will be quarantined and returned to Zimmer Sales Representatives to be destroyed. The public contact for the recall is: Consumer Relations Call Center Zimmer, Inc. P O Box 708 Warsaw, IN 46581-0708 [email protected] Phone: 1-800-447-5633 Fax: 574-371-8755
Quantity in commerce
83 units
Distribution
Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.
Product codes and lots
Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0853-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0853-2015Class IITerminatedVoluntary: Firm initiated