Z-0869-2015Class I
KimVent Microcuff Subglottic Suctioning, Endotracheal Tube, 9.0 mm, Distributed by Kimberly Clark Global...
Halyard Health / initiated 2014-11-17 / Terminated / root cause: manufacturing process
Halyard Health began a recall of KimVent Microcuff Subglottic Suctioning, Endotracheal Tube, 9.0 mm, Distributed by Kimberly Clark Global Sales, LLC. Used for airway management by oral... on . FDA classified it as Class I and gives the reason as "The cuff inflation line may detach from the endotracheal tube during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0869-2015
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Halyard Health
- Product
- KimVent Microcuff Subglottic Suctioning, Endotracheal Tube, 9.0 mm, Distributed by Kimberly Clark Global Sales, LLC. Used for airway management by oral intubation of the trachea and for removal of secretions that accumulate in the subglottic space.
- Product code
- BTR Tube, Tracheal (W/Wo Connector)
- Reason for recall
- The cuff inflation line may detach from the endotracheal tube during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Halyard Health, sent an "URGENT: VOLUNTARY PRODUCT RECALL" letter dated November 17, 2014 via FedEx to its Consignees/customers. The letter describes the product, problem and actions to be taken. 'the customers were instructed to please evaluate your facilities for unused inventory of KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes, if present, quarantine and discontinue use; complete the attached Distributor Product Recall Response Form via fax to: 678-254-0391 within five (5) business days of receipt of this letter; if you are in possession of the affected product lots, a Halyard Health representative will provide a return authorization and instructions on shipping product to a specified distribution center, and; follow your internal product recall procedures to identify and address affected product. If you require further assistance, please call Halyard Health Regulatory Affairs at 770-587-7400.
- Quantity in commerce
- 80 cases (10 tubes per case)
- Distribution
- Worldwide Distribution: US (nationwide) AL, AR, AZ, CA, CT, FL, GA, IN, KS, KY, MA, ME, MI, MO, MS, MT, ND, NH, OH, OK, OR, PA, TN, UT, VA, and WV; and countries of: Australia, Canada, Europe (i.e., France, Germany, Great Britain, Ireland, Italy, Netherlands, Portugal, Sweden, & Switzerland), Hong Kong, New Zealand, and United Arab Emirates.
- Product codes and lots
- Product Code: 13224, Lot numbers: AA3274, AA3274V02, AA4062, AA4062V05, AA4092V02, AA4104V01, AA4111V03, AA4174V02 & AA4190V01.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0865-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0866-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0867-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0868-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0869-2015 | Class I | Terminated | Voluntary: Firm initiated |