VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1...
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK began a recall of VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0 on . FDA classified it as Class II and gives the reason as "Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0821-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mitsubishi Heavy Industries, Ltd., Hiroshima Machinery Work
- Product
- VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result, ExacTrac may display a Digitally Reconstructed Radiograph (DRR) image taken from the angle for the 1st port image fusion, rather than the one taken from the angle for the intended port
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Mitsubishi sent a Field Safety Notice dated November 14, 2014, that was e-mailed to US end users on December 2, 2014, which provided actions to be taken by the end users. If users need any further information or support concerning this issue, they may contact MHI's Customer Service Representative, Brainlab AG in Germany, contact person: Franz Gum, Tel: +49-89-99-1568-0, Fax: +49-89-99-1568-33, E-mail: [email protected]. If any other abnormality is detected, stop operation and contact MHI's Customer Service representative immediately. --- An Updated Field Safety Notice (dated December 4, 2014,) was sent to end users to correct some terms used in the initial notification.
- Quantity in commerce
- Domestic: 4 units
- Distribution
- US Distribution to the states of NY, FL, TX and OH.
- Product codes and lots
- Serial Numbers: 201902, 203901, 203919, and 203924 (US only)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0821-2015 | Class II | Terminated | Voluntary: Firm initiated |