510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0823-2015Class II

Sirius Polish Cemented Stem 38C Catalog Number: 51-199335 Lot Number Identification: 141680, 180920 The...

Zimmer GmbH / initiated 2014-12-02 / Terminated / root cause: packaging

Biomet, Inc. began a recall of Sirius Polish Cemented Stem 38C Catalog Number: 51-199335 Lot Number Identification: 141680, 180920 The Biomet's Sirius Femoral Hip Stem is a double-tapered... on . FDA classified it as Class II and gives the reason as "After receiving a complaint, Biomet Orthopedic Inc., conducted an investigation which determined that the following parts were mixed up prior to packaging and shipment: Sirius Polish Cemented Stem 348 Catalog Number...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0823-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Sirius Polish Cemented Stem 38C Catalog Number: 51-199335 Lot Number Identification: 141680, 180920 The Biomet's Sirius Femoral Hip Stem is a double-tapered, highly polished cemented hip stem, designed for either primary or revision surgery. A distal centralizer is provided for optimal placement within the femoral canal. The system is used with Type 1 taper modular heads and compatible acetabular shells/liners and screws to provide a total hip replacement system.
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
After receiving a complaint, Biomet Orthopedic Inc., conducted an investigation which determined that the following parts were mixed up prior to packaging and shipment: Sirius Polish Cemented Stem 348 Catalog Number: 51-199333, Lot Number Identification: 024320 and Sirius Polish Cemented Stem 38C, Catalog Number: 51-199335, Lot Number Identification: 141680, 180920.
Root cause tag
packaging
FDA root cause
Mixed-up of materials/components
Action
Sent notices to customers dated 12/2/2014 notifying them of the recall: 5 were implanted, 3 were shipped to subsidiary for testing, and 2 are in Biomet control. Public Contact for the recall is: Audrey Daenzer, Field Action Specialist, Regulatory Compliance [email protected], can be contacted from Monday through Friday 8 AM to 5 PM, 574-372-1570.
Quantity in commerce
8
Distribution
Foreign Distribution in China and Australia only.
Product codes and lots
Sirius Polish Cemented Stem 38C Catalog Number: 51-199335 Lot Number Identification: 141680, 180920

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0823-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0822-2015Class IITerminatedVoluntary: Firm initiated
Z-0823-2015Class IITerminatedVoluntary: Firm initiated