Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on the... on . FDA classified it as Class II and gives the reason as "Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift observed up to 12% at MYO concentrations...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0946-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on the Dimension Vista System.
- Product code
- JIT Calibrator, Secondary
- Reason for recall
- Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift observed up to 12% at MYO concentrations within and above the reference range of the assay. Depending on quality control limits, this drift may not be detected.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- The firm, Siemens, sent an "Urgent Medical Device Recall"letter dated November 2014, to all customers who received the affected lot. The letter described the product, problem and actions to be taken. The customers were instructed to: Review this letter with your Medical Director; discard your remaining inventory; recalibrate with an alternate in-date lot of MYO CAL, and complete and return the attached Field Correction Effectiveness Check Form and fax the form to (312) 275-7795. Siemens will replace any unused inventory of the affected lot at no charge. If you have any questions, please contact your Siemens Customer Care Center - Technical Solutions at 800-441-9250 or your local Siemens technical support representative.
- Quantity in commerce
- 1230
- Distribution
- Worldwide Distribution: US (nationwide) and country of: Canada.
- Product codes and lots
- Lot 4FD085, exp. 12-01-2014
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0946-2015 | Class II | Terminated | Voluntary: Firm initiated |