BBL Port A Cul tubes, catalog number 221606. Patient sample collection tubes. BBL Port A Cul tubes, jars and...
Becton Dickinson & Co. began a recall of BBL Port A Cul tubes, catalog number 221606. Patient sample collection tubes. BBL Port A Cul tubes, jars and vials contain a reduced transport medium and are... on . FDA classified it as Class II and gives the reason as "The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the viability of anaerobic microorganisms". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1746-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BBL Port A Cul tubes, catalog number 221606. Patient sample collection tubes. BBL Port A Cul tubes, jars and vials contain a reduced transport medium and are intended to maintain the viability of anaerobic, facultative and aerobic microorganisms during transport from the patient to the laboratory. Sterile packages are for collection of specimens in clean areas; e.g., surgical suites.
- Product code
- N/A
- Reason for recall
- The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the viability of anaerobic microorganisms.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- BD Diagnostic Systems sent an Urgent Product Recall letter dated December 2014 to all affected customers. Recall notification was initiated on November 21, 2014, the notification was sent by fax and letters by UPS. Customer notification for the expansion of the recall was initiated on December 18, 2014, the notification was sent by fax and letters by UPS. For questions regarding this recall call 800-638-8663.
- Quantity in commerce
- 145,860 tubes
- Distribution
- US Nationwide distribution in AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
- Product codes and lots
- Part/Cat No. Lot / Serial No.Date of Mfr Exp. Date 221606 4042455 2/25/2014 2/5/2015 221606 4067180 3/18/2014 3/5/2015 221606 4087204 4/10/2014 3/25/2015 221606 4124702 5/8/2014 4/24/2015 221606 4161968 7/7/2014 6/12/2015 221606 4181959 7/14/2014 7/3/2015 221606 4032407 2/14/2014 1/29/2015 221606 4042069 2/17/2014 2/4/2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1746-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1747-2015 | Class II | Terminated | Voluntary: Firm initiated |