Z-1747-2015Class II
BBL Port A Cul tubes with Swabs Sterile Pack, catalog number 221607. For patient sample collection
Becton Dickinson & Co / initiated 2014-11-21 / Terminated / root cause: manufacturing process
Becton Dickinson & Co. began a recall of BBL Port A Cul tubes with Swabs Sterile Pack, catalog number 221607. For patient sample collection on . FDA classified it as Class II and gives the reason as "The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the viability of anaerobic microorganisms". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1747-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BBL Port A Cul tubes with Swabs Sterile Pack, catalog number 221607. For patient sample collection.
- Product code
- N/A
- Reason for recall
- The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the viability of anaerobic microorganisms.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- BD Diagnostic Systems sent an Urgent Product Recall letter dated December 2014 to all affected customers. Recall notification was initiated on November 21, 2014, the notification was sent by fax and letters by UPS. Customer notification for the expansion of the recall was initiated on December 18, 2014, the notification was sent by fax and letters by UPS. For questions regarding this recall call 800-638-8663.
- Quantity in commerce
- 92,710 units
- Distribution
- US Nationwide distribution in AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
- Product codes and lots
- Part/Cat No. Lot / Serial No.Date of Mfr Exp. Date 221607 4067181 3/19/2014 3/5/2015 221607 4083274 4/3/2014 3/20/2015 221607 4124706 5/9/2014 4/24/2015 221607 4127564 5/22/2014 5/1/2015 221607 4170891 7/8/2014 6/19/2015 221607 4006235 2/18/2014 12/26/2014 221607 4028420 2/19/2014 1/21/2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1746-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1747-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY