Z-0874-2015Class II

Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve...

Ebi, LLC / initiated 2014-11-11 / Terminated / root cause: design

Ebi, Llc began a recall of Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to... on . FDA classified it as Class II and gives the reason as "A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0874-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ebi, LLC
Product
Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.
Product code
OVE Intervertebral Fusion Device With Integrated Fixation, Cervical
Reason for recall
A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Biomet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated 11/11/2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately locate and remove from circulation the items consigned/loaned to your account as identified; carefully follow the instructions on the enclosed "Response Form"; complete and return Response Form via fax to 303.501.8444, and if you have further distributed these items to medical facilities, you MUST notify them of the actions. Questions related to this notice should be directed to the Sr. Quality Engineer at 518-859-3487 Monday through Friday, 8 am to 4 pm ET.
Quantity in commerce
193 units
Distribution
US Nationwide Distribution.
Product codes and lots
Part number 14-531593 Part number 14-531594  Catalog number Lot Number 14-531593 N23597  14-531594 N23599

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0874-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0874-2015Class IITerminatedVoluntary: Firm initiated