510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0825-2015Class II

Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular...

Depuy Orthopaedics Inc. / initiated 2014-11-07 / Terminated

DePuy Orthopaedics, Inc. began a recall of Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular surface of the femoral head in primary hip surgery and... on . FDA classified it as Class II and gives the reason as "Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0825-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular surface of the femoral head in primary hip surgery and for the salvage of a failed previous hip surgery.
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to tissue tension and not the taper connection between the femoral head and stem. The investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the
FDA root cause
Mixed-up of materials/components
Action
On 11/7/2014, URGENT INFORMATION - RECALL NOTICE notifications were sent to the affected distributors, hospitals, and doctors with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospitals and doctors were notified via sales consultant visit. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For clinical questions from surgeons, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT). For questions about device recall information provided, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT).
Quantity in commerce
19 units worldwide
Distribution
Distributed in the states of WV, AZ, and AL, and the countries of Germany, and Finland.
Product codes and lots
Lot 7928614 CATALOG NO. : 1365-36-740 Barcode GTIN: 10603295033684

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0825-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0825-2015Class IITerminatedVoluntary: Firm initiated