510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1016-2015Class II

Airo Mobile Intraoperative CT I Airo; Model #: MobiCT-32. Computed Tomography x-ray system

Brainlab, Inc. / initiated 2014-12-08 / Terminated / root cause: software

Mobius Imaging, LLC began a recall of Airo Mobile Intraoperative CT I Airo; Model #: MobiCT-32. Computed Tomography x-ray system on . FDA classified it as Class II and gives the reason as "There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1016-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Brainlab, Inc.
Product
Airo Mobile Intraoperative CT I Airo; Model #: MobiCT-32. Computed Tomography x-ray system.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect navigation data being transferred to the curve.
Root cause tag
software
FDA root cause
Software design
Action
Mobius sent an Urgent Field Safety Notice dated December 8, 2014, to all affected customers. The letter explains the recall and provides advise on action to be taken by the user. The letter also informs consignees that Mobius Imaging is working on a software fix for the software bug/communication error. The final fix is anticipated in early 2015. Advise on action to be taken by the user: The user should be prepared for the potential necessity to perform a manual navigation registration in order to use navigation. In general, do not use navigation data transmitted by the AIRO Mobile CT Scanner of any scan the Brainlab AIR App displayed the "Received Wrong Dataset" error for. To use such scans with the Curve, a manual navigation registration is required. To avoid the possibility of incorrect data from being transferred to the Curve, Mobius Imaging is recommending that customers only select "NO" when they come to the "Continue with same patient" screen after each scan. They can select the same patient under the "Existing Patients" tab and restart the workflow. Additionally, always adequately verify the accuracy of the navigation registration data as required by the Brainlab Curve" Navigation system software and the AERO Operator's Manual. If you require further clarification, please feel free to contact your local Brain lab Customer Support Representative. Customer Hotline: +49 89 99 15 68 44 or +1800 597 5911 (for US customers) or by E-mail: [email protected] (for US customers: [email protected]) Fax Brainlab AG: + 49 89 99 15 68 33 Address: Brainlab AG (headquarters), Kapellenstrasse 12, 85622 Feldkirchen, Germany. For questions regarding this recall call 978-615-5025.
Quantity in commerce
21 devices distributed. 15 of these devices affected by the software v. 1.1.1+ patch
Distribution
USA including AK, IL, WA, MA, NY, WV, NC, TX, CA, PA and Internationally to Germany, Netherlands, Belgium, Switzerland, Saudi Arabia, Egypt.
Product codes and lots
Serial #s: 2251 (#1)*, 2324 (#2), 2328(#3), 2344 (#4), 2346 (#5), AIRO-0106, AIRO-0107*, AIRO-0108, AIRO-0110, AIRO-0111, AIRO-0112, AIRO-0113, AIRO-0114, AIRO-0115, AIRO-0116, AIRO-0117, AIRO-0118, AIRO-0119, AIRO-0120, AIRO-0122, and AIRO-0123. *Demo system

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1016-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1016-2015Class IITerminatedVoluntary: Firm initiated