Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0951-2015Class II

Labeled in part: I Color Complete / Crazy Lens, blister packing, 14.5mm, (Additional colors include: #003...

Intertrade Imports Inc. / initiated 2014-09-10 / Terminated / root cause: manufacturing process

Intertrade Imports Inc. began a recall of Labeled in part: I Color Complete / Crazy Lens, blister packing, 14.5mm, (Additional colors include: #003 White out, #005 Yellow out, #020 Red out, #014 Black... on . FDA classified it as Class II and gives the reason as "A surveillance sample collected by FDA confirmed the presence of Bacillus Cereus and Bacillus Lentus in the contact lenses, product is not sterile. I Color Color Complete brand color contact lenses are unapproved...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0951-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intertrade Imports Inc.
Product
Labeled in part: I Color Complete / Crazy Lens, blister packing, 14.5mm, (Additional colors include: #003 White out, #005 Yellow out, #020 Red out, #014 Black out, #070 Manson, #136 Lava, #009 Yellow cat, #073 Red cat, #050 White cat, #032 Soccer ball, #077 Spiral, #071 Target, #057 Wild fire, #056 Spider web, #174 Black moon, #179 Green moon, #180 Zombie, #181 Green hole, and #184 Black hole among others styles) Product Usage - Eye contact lens
Product code
LPM Lenses, Soft Contact, Extended Wear
Reason for recall
A surveillance sample collected by FDA confirmed the presence of Bacillus Cereus and Bacillus Lentus in the contact lenses, product is not sterile. I Color Color Complete brand color contact lenses are unapproved medical devices.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
The firm verbally contacted each distributor on September 10, 2014 to notify them about the recall and actions that needs to be taken.
Quantity in commerce
2850 each
Distribution
Worlldwide Distribution - USA including Texas, California, Florida, and St. Lucia, Barbados and St. Marten.
Product codes and lots
ALL LOTS / CODES

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0951-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0950-2015Class IITerminatedVoluntary: Firm initiated
Z-0951-2015Class IITerminatedVoluntary: Firm initiated