Z-0953-2015Class II

The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The...

Depuy Orthopaedics Inc. / initiated 2014-11-10 / Terminated / root cause: design

DePuy Orthopaedics, Inc. began a recall of The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The instrument interfaces with several related ATTUNE... on . FDA classified it as Class II and gives the reason as "Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0953-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The instrument interfaces with several related ATTUNE instruments: keel punch, tibial tower, evaluation bullet, fixed bearing tibial impactor, rotating platform tibial impactor, and femoral impactor, for the purpose of impaction, extraction, and insertion applications at steps within the surgical procedure. Due to the frequency of its use, two impaction handles are supplied in each kit.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.
Root cause tag
design
FDA root cause
Device Design
Action
On 11/10/2014, URGENT INFORMATION ATTUNE INTUITION IMPACTION HANDLE AND IMPACTORS DEVICE CORRECTION NOTICE notifications were sent to the affected distributors and medical professionals. All distributors were notified via electronic mail, and all medical professionals with affected inventory were also notified by distributor or Sales Consultant via in person delivery or mail or electronic mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT).
Quantity in commerce
7,944
Distribution
Worldwide Distribution-US (nationwide) except AK and RI, including PR and the countries of Chile, Canada, Australia, Austria, Belgium, France, Hong Kong, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK
Product codes and lots
CATALOG NO. : 2544-01-017 LOT NUMBER: All Lots Barcode GTIN: 10603295423324

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0953-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0953-2015Class IITerminatedVoluntary: Firm initiated