Z-0972-2015Class II

Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is dedicated...

Siemens Medical Solutions USA, Inc. / initiated 2014-12-03 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is dedicated angiography systems developed for single and biplane... on . FDA classified it as Class II and gives the reason as "Suboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0972-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Suboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure could be terminated in rare cases.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent an important customer safety information notice, dated December 3, 2014, to all affected customers. The letter identified the product, problem, and action to be taken. The following hardware modifications will be implemented with the corrective action AX042/14/S: 1. A modified cable outlet to provide additional space for the cabling. 2. A modified plastic cover at the cable outlet providing optimal cable guidance in the area of the arc. In general all systems will be checked for existing damage prior to the implementation of this action. For questions regarding this recall call 610-219-6300.
Quantity in commerce
50
Distribution
Nationwide Distribution
Product codes and lots
model numbers: 10848281; 10848354

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0972-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0972-2015Class IITerminatedVoluntary: Firm initiated