Z-0994-2015Class II

Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio...

Siemens Medical Solutions USA, Inc. / initiated 2014-12-15 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio Max is a radiographic system used in hospitals... on . FDA classified it as Class II and gives the reason as "Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0994-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio Max is a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio Max system is not meant for mammography. The Ysio Max uses integrated or portable digital detectors for generating diagnostic images by converting x-rays into electronic signals. Ysio Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Siemens, sent a "Safety Advisory Notice - XP080/14/S" letter dated December 8, 2014, to end users that identified the product, problem, and actions to be taken. Siemens provided instructions to avoid the issue and indicated a software fix would be released as a permanent resolution. The customers were instructed observe this safety notice and comply with the corresponding measure until the update has been fully completed. If you have any questions, contact the Regulatory Technical Specialist at 610-219-2119.
Quantity in commerce
11
Distribution
US Distribution to states of: ID, IL, MI, MN, MO, ND and PA.
Product codes and lots
Material 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470  serial numbers24077 24051 24038 24059 24066 24015 24026 24058 24039 24090 24075

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0994-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0994-2015Class IITerminatedVoluntary: Firm initiated