Z-1026-2015Class II

Beckman Coulter, Lactate Assay, Catalog No. OSR6X93

Siemens Healthcare Diagnostics Inc. / initiated 2014-12-15 / Terminated

Beckman Coulter Inc. began a recall of Beckman Coulter, Lactate Assay, Catalog No. OSR6X93 on . FDA classified it as Class II and gives the reason as "N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1026-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter, Lactate Assay, Catalog No. OSR6X93
Product code
KHP Acid, Lactic, Enzymatic Method
Reason for recall
N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations for the treatment of acetaminophen overdose, interferes with the assays for these analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. NAC interference may lead to falsely low results.
FDA root cause
Under Investigation by firm
Action
A customer notification letter was sent to their customers on 12/15/14 to inform them that Beckman Coulter has recently confirmed that N-Acetyl Cysteine (NAC), when administered in therapeutic concentrations (for the treatment of acetaminophen overdose), has been determined to interfere with assays for the following analytes: Cholesterol OSR6X16, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to call (800) 223-0130 in the US and Canada. Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.
Quantity in commerce
31,799 units total (15,841 units in US)
Distribution
Nationwide in US and Puerto Rico. World Wide: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Finland, France, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guam, Honduras, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Lebanon, Lesotho, Liberia, Libyan Arab Jamahiriya, Lithuania, Malawi, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam, Zambia, and Zimbabwe
Product codes and lots
All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1026-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1023-2015Class IITerminatedVoluntary: Firm initiated
Z-1024-2015Class IITerminatedVoluntary: Firm initiated
Z-1025-2015Class IITerminatedVoluntary: Firm initiated
Z-1026-2015Class IITerminatedVoluntary: Firm initiated