Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1081-2015Class III

VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813...

Lacrimedics, Inc. / initiated 2014-11-26 / Terminated / root cause: manufacturing process

Lacrimedics Inc began a recall of VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813, 1814, 1815, 1903, 1904, and 1905. This device is... on . FDA classified it as Class III and gives the reason as "Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1081-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Lacrimedics, Inc.
Product
VisiPlugST for the Lacrimal Efficiency Test. It is also sold in the EU as CollaSyn Plugs. Model number 1813, 1814, 1815, 1903, 1904, and 1905. This device is cylindrical in shape, approximately 1.75-2.00mm in length, and available in three sizes: 0.3mm, 0.4mm and 0.5mm. Plugs are place in the upper and/or lower horizontal canaliculus. Two plugs are placed into a foam carrier and nitrogen sealed inside a foil pouch. The foil pouch is sent to sterilizer for gamma sterilization. Two sterile pouches are packaged in a box with an IFU.
Product code
LZU Plug, Punctum
Reason for recall
Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact. The firm's data also suggests that a 4 year shelf life may not be sustained.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm notified their foreign consignees on 11/26/14. The firm made a decision to conduct a domestic recall on 12/4/14 and will notify US consignees Consignees will be provided with a return label and request any devices that they still have be returned to the firm. Any consignees that have gone out of business will be sent a certified letter to their last known address. CORRECTIVE ACTIONS: The firm has halted sales of the device until the drying method can be improved. Any remaining stock has been placed in quarantine and will be scrapped.
Quantity in commerce
781 units
Distribution
Worldwide Distribution-US (nationwide) including the states of AR, CA, FL, ID, IL, KS, MA, MI, NC, NJ, NY, OR, PA, and TX and to the countries of Australia, Brazil, Czech Republic, France, Ireland, Mexico, Peru, Switzerland, and United Kingdom.
Product codes and lots
LOT numbers:  072314-2328;  072314-2329;  072314-2330;  081314-2365;  081314-2366;  081314-2367;  091814-2345; 091814-2350;  100614-2344;   EXPIRATION DATES:  2018-05-15; 2018-07-23; 2018-08-13; 2018-08-14; 2018-09-14; 2018-09-14; 2018-09-25; 2018-10-06;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1081-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1081-2015Class IIITerminatedVoluntary: Firm initiated