Z-0908-2015Class II
I.M. Hole Locator Instrument. Orthopedic manual surgical instrument. The IM Hole locator guides the IM drill...
Depuy Orthopaedics Inc. / initiated 2014-11-12 / Terminated / root cause: design
DePuy Orthopaedics, Inc. began a recall of I.M. Hole Locator Instrument. Orthopedic manual surgical instrument. The IM Hole locator guides the IM drill during IM Hole preparation. The instrument is... on . FDA classified it as Class II and gives the reason as "Use of excessive force when impacting may lead to intra-operative femoral fracture". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0908-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- I.M. Hole Locator Instrument. Orthopedic manual surgical instrument. The IM Hole locator guides the IM drill during IM Hole preparation. The instrument is positioned between the condyles and aligned with the anatomical axis. Pins can be placed in the IM hole locator to assist with additional stability. The drill bush may also be lightly impacted onto the distal femur for additional stability.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Use of excessive force when impacting may lead to intra-operative femoral fracture.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, DePuy Synthes, sent an "URGENT INFORMATION - DEVICE CORRECTION NOTICE HP LCS IM Hole Locator" notification dated November 12, 2014 to the affected distributors. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The distributors were instructed to: review the instructions and the information in this notice; provide this notice to your sales consultants, and complete and return the DISTRIBUTOR RESPONSE CARD: Verify Sales Consultant Notifications via fax to (574) 371-4939 or email: [email protected]. For questions, please contact Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EST).
- Quantity in commerce
- 1130
- Distribution
- Worldwide Distribution: US (nationwide) including states of: AZ, CA, CO, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MT, NC, NV, NY, OH, PA, TN, TX, VA, WA, and WI; and countries of: Canada, Australia, Austria, Belgium, Czech Republic, France, Germany, India, Israel, Italy, Japan, Malaysia, Russia, South Africa, Switzerland, and United Kingdom.
- Product codes and lots
- CATALOG NO. : 9505-01-041 Barcode GTIN: 10603295224389 ALL LOTS
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0908-2015 | Class II | Terminated | Voluntary: Firm initiated |