Z-1017-2015Class I

Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and...

Medtronic, Inc. / initiated 2014-12-10 / Terminated / root cause: labeling and UDI

Covidien began a recall of Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended for... on . FDA classified it as Class I and gives the reason as "A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1017-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien Trellis-6 Peripheral Infusion System. Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01. Sterile EO. Product Usage: Intended for controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
Product code
KRA Catheter, Continuous Flush
Reason for recall
A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
Consignees were sent an Urgent Product Recall letter dated 12/15/2014. The letter described the issue, identified the affected product, and the required actions. Affected product was to be returned directly to customers' Covidien Sales Rep; or customers were to contact Covidien Service at 1-800-716-6700 to arrange for product return. Questions can be directed to the Covidien Sales Rep or to Covidien Service at 1-800-716-6700. Customers were to complete and return the Verification Form along with the unused product as soon as possible.
Quantity in commerce
216 devices (207 US, 9 OUS)
Distribution
Worldwide Distribution - US Nationwide and the countries Australia, Canada, Finland, France, Germany, Ireland, Italy, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Product codes and lots
Model BVT608010, Lot numbers: 9853174, 9864497, 9890772, 9890931, 9925676, 9932672, 9937308, 9937315, 9940474, 9940614, 9976340, 9978233, A001589, A001740.   Model BVT608030, Lot Numbers: 9853331, 9887597, 9887695, 9887757, 9932042.   Model BVT612010, Lot numbers: 9854121, 9868214, 9868311, 9875472, 9875632, 9876174, 9883295, 9884788, 9884951, 9886196, 9886312, 9932096, 9968511, 9974109, 9974126, 9976472, 9977091, 9978862, A002299, A002423.   Model BVT312030, Lot numbers: 9854124, 9892081, 9941213, 9971096, 9971139, 9976506, 9977140, 9978964, 9979548, A003021, A003121, A003651, A007900, A008492, A008603, A008687.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1017-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1017-2015Class ITerminatedVoluntary: Firm initiated
Z-1018-2015Class ITerminatedVoluntary: Firm initiated