Baxter Integrated APD Set with Cassette 3-Prong, Sterile, nonpyrogenic fluid path
Baxter Healthcare Corp. began a recall of Baxter Integrated APD Set with Cassette 3-Prong, Sterile, nonpyrogenic fluid path. Product Usage: The recall device is indicated for use in the treatment of... on . FDA classified it as Class II and gives the reason as "Baxter is issuing a recall for one lot of Integrated Automated Peritoneal Dialysis set with Cassette 3 prong due to complaints received for leakage". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1039-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Integrated APD Set with Cassette 3-Prong, Sterile, nonpyrogenic fluid path. Product Usage: The recall device is indicated for use in the treatment of patients with renal failure to provide PD fluid exchanges with Baxter HomeChoice PD Systems in clinical and home use settings.
- Product code
- FKX System, Peritoneal, Automatic Delivery
- Reason for recall
- Baxter is issuing a recall for one lot of Integrated Automated Peritoneal Dialysis set with Cassette 3 prong due to complaints received for leakage.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Baxter sent an Urgent Product Recall notification letter dated January 8, 2015 . The letter identified the affected product, problem and action to be taken. The recalling firm has asked the consignees to 1) locate and remove all affected product from their facilities, 2) contact Baxter to arrange for return and credit, and 3) complete the attached customer reply form and return it to Baxter. For questions call 800-422-9637.
- Quantity in commerce
- 8,640 units
- Distribution
- US Nationwide Distribution in the states of FL, WI, AR, TN, TX, NC, MO, NM, NJ, CT, PA, NV, NY, CA, MI, WI, KS, and OH.
- Product codes and lots
- H13L22013
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1039-2015 | Class II | Terminated | Voluntary: Firm initiated |