ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The...
Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution... on . FDA classified it as Class II and gives the reason as "Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0987-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.
- FDA root cause
- Under Investigation by firm
- Action
- A Siemens service provider visited each site to verify and/or update customer configuration on the ADVIA Chemistry XPT and CentraLink Systems to ensure proper reporting of results. An Urgent Medical Device Correction letter was issued on December 16, 2014 to all affected customers. Customers were also provided a Customer Bulletin providing instructions on how configuration setting should remain.
- Quantity in commerce
- 7
- Distribution
- Distributed in the state of WA.
- Product codes and lots
- CentraLink" Data Management System V14x Catalog Numbers: 10817209, 10818262
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0987-2015 | Class II | Terminated | Voluntary: Firm initiated |