Z-0987-2015Class II

ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The...

Siemens Healthcare Diagnostics Inc. / initiated 2014-12-16 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution... on . FDA classified it as Class II and gives the reason as "Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0987-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.
FDA root cause
Under Investigation by firm
Action
A Siemens service provider visited each site to verify and/or update customer configuration on the ADVIA Chemistry XPT and CentraLink Systems to ensure proper reporting of results. An Urgent Medical Device Correction letter was issued on December 16, 2014 to all affected customers. Customers were also provided a Customer Bulletin providing instructions on how configuration setting should remain.
Quantity in commerce
7
Distribution
Distributed in the state of WA.
Product codes and lots
CentraLink" Data Management System V14x Catalog Numbers: 10817209, 10818262

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0987-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0987-2015Class IITerminatedVoluntary: Firm initiated